Validazione del Bio-Explorer
A NON-INVASIVE BIOPOTENTIAL ELECTRODE FOR THE CORRECT DETECTION OF BIOELECTRICAL CURRENTS
C. Boccaletti1,
F.Castrica2, G. Fabbri1, M. Santello3
1 Department of Electrical Engineering – “Sapienza” University of
Rome
Via Eudossiana 18, 00184 Rome, Italy
2 Biophysics Research srl – Via di Quarto Rubbie, 45 00040 Rome Italy
e-mail: gianluca.fabbri@uniroma1.it
3 Department of Kinesiology – Arizona State University
PEBE 107B, Orange Street, Tempe, AZ
85287-0404, USA
Abstract
An advanced current measure probe is
described for biological applications. Singular points on the body skin emit a
semi-constant current that could be potentially useful for a medical diagnosis
of the body organs to which any singular point is connected. In order to achieve
an accurate and reliable analysis of the emitted current, all the error sources
have been studied in order to design a ready to use probe.
Introduction
Significant effort has been devoted to
evaluate the electrical properties of paths on the human body described by
Chinese Medicine (i.e., meridians) [1]. The specific points of interest
(acupuncture points) are described as being one millimeter in diameter. The
electrical characteristics of the acupuncture meridians have been extensively
studied since the 1950s to establish their existence by scientific methods
[2]-[3]. To date, the bioelectric properties of acupuncture points are not
clearly understood. For example, many discrepancies and inconsistencies appear
in the literature concerning skin impedance on the acupuncture points
[4].The
change in impedance at acupuncture points is reportedly anywhere from 1/2 to
1/20 of the impedance of the surrounding skin. These low impedance points are
thought to be the result of sensory and motor nerves emerging from deep tissue
to superficial layers of the skin (see fig.1)[5]. It is by now generally
accepted that both meridian and acupuncture points have lower electrical
resistance or impedance than nearby surrounding points. Small electrical
potentials can be recorded on the skin at the terminal points of acupuntcure
meridians and the results are replicable under controlled experimental
conditions [6]-[7].
As a results of these potentials, the current produced
ranges from a few to several hundred nanoamperes, which can be recorded by
an external measuring device. The electromotive force producing these endogenous
currents results from electrical potentials at the acupuncture points on the
skin and is a function of the internal resistance of the point.

Figure 1 An acupuncture point seen as
the result of sensory and motor nerves emerging from deep tissue to superficial
layers of the skin
Figure 2 shows a typical
behaviour of the electric resistance on an acupuncture point. The resistance is
a function of the cutaneous region, skin humidity and pressure of the
measurement electrode. The arrow points out the strong decrease of resistance in
the point of acupuncture. The ordinate reflects the resistance in ohms, the
abscissa the run of the electrode on the cutaneous surface. Most of research on
electroacupuncture has been based on skin resistance and feeding in external
currents [8]-[9].
All of these instruments use an external voltage source to
measure the skin conductance and the measurement can be easily influenced and
disrupted by many factors, making it difficult to achieve reliable and
replicable results. Moreover, feeding the point with an
external current in order to measure the internal resistance constitutes an
interference from a physiological point of view and the exact effect on the
human organism of applying external invasive currents of even a few
μA to acupuncture points are not know. Thus,
any method permitting readings to be taken in nanoamperes at acupuncture points
without feeding in addictional external current would be diagnostically useful.
Figure 3 shows the schematic representation of current measurement without use
of an external generator[10].
Figure 2 Electrical resistance of the acupuncture point
A pointed electrode placed on an
acupuncture point acquires the current and a ground reference electrode with a
large surface area rests on the subject’s forearm. This measurement is able to
detect the current emitted by the acupuncture point when it is short circuited
on the instrument, but some experiences has pointed out the scarce reliability
and inadequacy of this kind of measurement technique due to unavoidable error
components.
Figure 3 Schematic
representation of current measurement without use of an external voltage source.
The main purpose of this
study was to improve such measurement technique using a new non-ivasive
biopotential electrode designed to solve most of the above measurement problems.
Our experience has confirmed the presence of electrical potentials of several
millivolts (between few mV and 500 mV) that can be detected on the surface of
the skin at acupuncture points. The body’s conductivity determines the amount of
current produced by electromotive force resulting from differences in these tiny
potentials and if current can be measured in a replicable manner, free of
external interference, the result could be potentially useful for medical
diagnostics. Considering the internal resistance of the points, the generated
short circuit dc currents are expected to fall in the range of 10nA to 5000 nA.
Main error components
For the correct evaluation of the
physiological parameters of medical interest it is necessary that the measures
are repeatable and reliable. We performed an analysis of the main error
components that may influence the measurements:
-
1. It is necessary to avoid the formation of
spurious potentials that would contaminate our measurements. In particular, the
measurement technique must not depend on the skin ph and on the internal
resistance of the characteristic point. This can be achieved using electrodes
that minimize the overpotentials, such as the silver/silver chloride ones.
-
2. The tip of the probe must have a proper shape in order to minimize pain for the
patient and to keep a costant contact area on the skin. The contact area will
vary as a function of the pression exerted by the operator whereas the measured
current will be a function of pressure. Therefore a pressure compensation system
must be introduced such that measurements are virtually insensitive to
fluctuations in pressure over a wide range of values.
-
3. The current emitted by the characteristic point, when measured in a short
circuit, decreases asymptotically from a peak value to a lower steady state
value as a function of many physiological parameters and conditions. Measured
data validation is therefore demanded to a software procedure to calculate the
standard deviation of the datum in a given number of samples. When the standard
deviation is lower of a preset threshold, the current value is accepted and
displayed.
The silver/silver-chloride electrode
To measure and record potentials and
currents in the body an electrode can be used as biomedical sensor [11]. This
seems to be a very simple function, but in fact an electrode recording
biopotentials is actually a transducer, converting ionic currents in the body
into electronic currents in the electrode. This transduction function greatly
complicates electrode design. Figure 4 shows an electrode-electrolyte interface.
The electrode only has one type of charge carrier (electron), whereas the
electrolyte has two types of charge carriers (cation and anion). The electrolyte
is an aqueous solution containing cations of the electrode metal C+ and anions
A-. The electrode consists of metallic atoms C and the current crosses the
interface from left to right. At the interface, charge is exchanged through
chemical reactions, which can be generally represented as:
C⇔Cn+ +ne-
Am-⇔A+me-
where n is the valence of C and m is the valence of A.
Figure 4 Electrode-electrolyte interface
A potential difference know as the
half-cell potential is determined by the metal involved, the concentration of
its ions in solution, and the temperature. The standard half cell potential, E0,
is the potential for 1M concentration solution at 25°C when no current flows
across the interface. When a circuit is constructed to allow current to flow
across an electrode-electrolyte interface, the observed half-cell potential is
often altered. The difference between the observed half-cell potential for a
particular circuit and the standard half cell potential is known as the
overpotential. Three basic mechanisms contribute to the overpotential:
ohmic, concentration, and activation.
-
1. The ohmic overpotential is the voltage drop across the electrolyte itself due to the
finite resistivity of the solution. Overall, this is usually not a big voltage
in high concentration solutions.
-
2. The concentration
overpotential results from changes in ionic concentration near the
electrode-electrolyte interface when current flows. Oxidation-reduction reaction
rates at the interface change with excess charge due to a finite current. This
modifies the equilibrium concentration of ions changing the half-cell potential.
-
3. Charge transfer in the oxidation-reduction reaction at the interface is not
entirely reversible. For metal ions to be oxidized, they must overcome an energy
barrier. If the direction of current flow is one way, then either oxidation or
reduction dominates, and the height of the barrier changes. This energy
difference produces a voltage between the electrode and the electrolyte, known
as the activation overpotential.
The overpotential of an electrode is then given by the sum of these three polarization mechanisms:
Vp=Vr+Vc+Va
where Vr is the ohmic overpotential,
Vc is the concentration overpotential, and Va is the activation overpotential.
Note that overpotentials impede current flow across the interface. A way to
minimize Vp is to use nonpolarizable electrodes. These allow conduction current
to flow across the interface with no energy exchange and there are no
overpotentials for this type of electrode. The best electrode to use for all
possibilities for biological electrode system is the silver/silver chloride
(Ag/AgCl) electrode. This is made of a silver metal base with attached insulated
lead wire coated with a layer of the ionic compound AgCl. The electrode is then
immersed in an electrolyte bath in which the principle anion of the electrolyte
is Cl-. For the best results, the electrolyte solution should also be saturated
with AgCl so there is little chance for any of the surface film on the electrode
to dissolve. Cl- is an attractive anion for electrode applications with mammals
since these animals (including humans) have an excess of chloride ions in
solution. The electrode-electrolyte interaction is described by the reaction
Ag⇔ Ag++e-
Ag++Cl-⇔ AgCl(Precipitate)
And the Nerst equation for the reaction can be written as:
RT RT
E=E0+——In(Ks)-——In(a α-)
nF nF
The first two terms on the right side
of this last expression are constants - only the third is related to ionic
activity. In biological systems, the large chlorine ion concentration makes its
activity fairly constant. This means that the half-cell potential for this
electrode is quite stable for biological systems. In this article we will only
consider low current densities, and consequently the electrode-electrolyte
interface can be modeled as a linear system with an equivalent circuit composed
exclusively of linear components (i.e., voltage/current sources, resistors,
capacitors and inductors). The terminal characteristics of an electrode have
both resistive and reactive components.
Figure 5 Equivalent circuit for a biopotential electrode in contact with an
electrolyte. Ehc is the half-cell potential, R
d and C
d make up
the impedance associated with the electrode-electrolyte
interface and polarization effects, and Rs
is the series resistance associated with interface effects and due to
resistance in the electrolyte
Fig. 5 shows the equivalent circuit of
the electrode-electrolyte interface. In this circuit Rd and Cd represent the
resistance (i.e., conduction currents) and the capacitance (i.e., displacement
currents) respectively resulting from the double-layer of ionic charge at the
electrode-electrolyte interface. The resistance Rs is the series resistance
associated with equivalent losses in the electrolyte itself.
There are three primary
layers in the skin. The outermost layer, or
epidermis, plays the most important role in the electrode-skin interface. It is
a constantly changing layer, the outer surface of which consists of dead
material on the skin’s surface with different electrical characteristics from
live tissue. The deeper layers of skin contain the vascular and nervous
components of the skin as well as the sweat glands, ducts, and hair follicles.
These layers are similar to others in the body, and with the exception of the
sweat glands, can be modeled as equivalent to the electrical characteristics of
the rest of the viscera. Given this anatomy, a general equivalent circuit
describing the characteristics of both the electrode-electrolyte interaction and
the connection to the skin can be developed, as illustrated in figure 6.
Figure 6 Total electrical
equivalent circuit obtained for a
body-surface electrode placed against skin
The epidermis can be
considered a semipermeable membrane to ions, so a potential given by the Nernst
equation, can be developed if there is a difference in ionic concentrations
across this membrane. The dermis and subcutaneous layer under it behave in
general as pure resistances. They generate negligible DC potentials. Finally,
the electrical characteristics of the sweat glands must also be taken into
account for a complete model of a skin electrode. The fluid secreted by sweat
glands contains Na+, K+, and Cl- ions, the concentrations of which differ from
those in extracellular fluid. This produces a potential between the lumen of the
sweat duct and the dermis and subcutaneous layers. There is also a parallel RpCp
combination with this potential representing the wall of the sweat gland and
duct. This equivalent model has been used by
many authors to simulate the electrical behaviour of the electrode-skin
interaction.
The Non-invasive Probe and the pressure compensation system
To realize the measuraments, a
noninvasive probe has been designed. The probe is illustrated in Fig.7 and its
electrode is a rod 2 mm in diameter made of casting of Ag-AgCl powder. The rod
must minimize the skin pain to the patient during the measurements; to achieve
this goal it has been decided to use a semisphere shaped rod instead of a
cylindrical one because the last one could injure the skin at some probe-skin
angles.
Figure 7 The realized probe
The contact surface will increase or
decrease according to the pressure and pressure unbalance will introduce an
error at a rate proportional to the unbalance itself. Also the contact
resistance will be influenced by the pressure. It can be easily seen that an
error of more than 100% of the measured current can be introduced for a pressure
variation of about 300 grams. Assuming that the reliability of the measure must
be as high as possible, the above mentioned error cannot be tolerated, hence a
pressure compensation system must be introduced to make the measure almost
insensitive to pressure changes for a wide span of values. It was then studied a
pressure compensation system to allow the operator to push the probe on the
measure point to a well defined and repetitive pressure.
Figure 8 Probe dimensions
The presence of a pressure sensor
allows the operator to work in a linear range of pressures, namely between 100
and 250 grams, in order to get the contact surface variation in a linear range.
The probe tip holder can slide in its guide for about 1mm to transmit the force
on the pressure gauge, but at the same time, a perfect guide sealing must be
provided in order to avoid the saline solution to penetrate the inner space of
the housing of the probe. A sealing grease should have in any case very low
friction, in order to assure a complete repositioning [reset] of the tip after
measures. The residual pressure on the sensor must be at any rate checked by the
application software before any current measure. Figure 8 shows the probe
dimensions.
Data validation
As seen in thousands of measures, the
current emitted by acupuncture points of the body, when measured in a short
circuit, decreases asymptotically from a peak value to a lower steady state
value, as a function of many physiological parameters and conditions. Measured
data validation is therefore demanded to a software procedure. A software has
been then developed to calculate the standard deviation of the data in a given
number of samples. When the standard deviation is lower of a preset threshold,
the current value is accepted and displayed. Many characteristics points on
different subjects have been tested and some results are reported in the
following paragraph.
Figure 9 Current deviation
As an example, figure 9 shows some
results obtained measuring for ten times the current emitted by two
characteristics points (point A and point B) of a male subject. Measurement were
taken every 30 seconds with a nominal pressure of 200 gr. Point A is the point
LU11(Shao Shang) of the Lung Channel and is located on the radial side of the
thenar eminence, 0.15 cm posterior to the nailbed. Point B is the point HE9
(Shao Chong) of the Heart Channel and is located on the radial side of the fifth
digit, 0.15 cm from the corner of the nail bed. Figure 10 shows the locations of
the two points on the hand of the subject and the locations of the two channels
along the arm.
Figure 10 Location of point A and B
Results
The probe was used to test some chosen
characteristics points. The currents were recorded using BE 100-04 apparatus
(patented by Biophysics Research srl, Rome, Italy), set for the appropriate
measuring mode. All the measurements were done holding the probe in an upright
position, at an angle of about 80 degrees with the skin, with the tip in contact
with the acupuncture point to be tested. The majority of tests took place during
the day between 9am and 6pm. Table 1 and Figure 11 show the results of measuring
for 8 times the currents emitted by Point A at different moments during the day.
All the indicated currents are in nA and the table also provides the standar
deviation (SD) and the mean value (MV). Table II shows the results of measuring
the currents emitted by Point A in 12 different subjects for 10 times with a
time step of 1 minute. Figure 12 illustrates the measurements on the 12
subjects.
Table 1 Different current measurements during the day
Figure 11 Different current measurements during the day
Table 2 Different current measurements in 12 subjects
Figure 12 Measurements of current emitted by Point A in 12 different subjects
Conclusion
This study shows that small electric
currents can be measured at characteristic points on the skin in a replicable
manner with a non-invasive probe. A correct evaluation of the magnitude of this
currents could be significative for a diagnosis of the organs correlated with
the points. More investigation need to be done in order to understand the
correlation between any measure point and the activity of the body organs.
REFERENCES
[1]
M. Reichmanis, A. Andrew, R.O. Becker RO, Electrical correlates of acupuncture points,
IEEE Trans Biomed Eng, 1975,
22:533-235.
[2]J.J. Tsuei, The
science of acupuncture ,
IEEE Engineering in medicine and biology,
May-June 1996 52-57
[3]K.G. Chen, Electrical
properties of meridians,
IEEE Engineering in medicine and biology, 15(3)1996, 58-63
[4]G. D. Mc Carrol, B.
A. Rowley, An Investigation of the Existence of Electrically Located Acupuncture
Points,
IEEE Transactions on
Biomedical Engineering,Vol.BME-26, NO. 3,
March 1979
[5]
H. Heine, Manuale di Medicina
Biologica, 207-215, Guna Editore, 1999, Milano
[6]Z. X. Zhu, Research
advances in the electrical specificity of meridians and acupuncture points”,
Am J Acupunct,
(9)1981, pp. 203- 216.
[7]K.G. Chen,
Biological Implication of Electrical Properties of Acupuncture Meridians,
Proceedings of the 20th Annual International Conference of the
IEEE Engineering in Medicine and Biology Society,
Vol. 20, No 2,1998, 1086-1087.
[8]E. F. Prokhorov, J.
G. Hernandez, Y. V. Vorobiev,E. M. Sinchez, T. E. Prokhorova, G. Zaldivar,
Electrophysical Characterization of Biologically Active Points and Human Skin by
in Vivo Impedance Measurement,
Proceedings of the 22"d Annual EMBS International Conference, July 23-28,2000,
Chicago IL.
[9]
W.A. Lu, J.J. Tsuei,
K.G. Chen, Preferential direction and symmetry of electric conduction of human
meridians,
IEEE Eng. Med. Biol. Mag., 1999 vol. 18, p. 76.
[10]
R. Eickhorn,
H.Schimmel, Electrophysiological diagnosis at terminal poits of acupuncture
meridians,
Biomedical Therapy,
vol XVII, 1999 n.3.
[11]
M. O’Donnell,
Biomedical Instrumentation & Design, BME/EECS 458, Chapter 3,
Medical Instrumentation Application and Design
(3rd Edition) John G. Webster Editor